X-ray system configuration:

  • system based on floor-mounted
  • X-ray diagnostics table
  • X-ray stand
  • X-ray high-voltage generator (power supply)
  • X-ray emitter (generator)
  • Flat Panel Detector in the table and in the stand
  • Digital system of image processing

X-Ray Diagnostics table

  • Size of the table surface - not less than 2010 х 820 mm
  • Adjusting the height of the table above the floor - In the range of 500 to 800 mm
  • Table load - not less than 220 kg
  • Table side-to-side movement range - not less than 1000 mm
  • The range of flat detector motorized lengthwise motion - not less than 500 mm
  • Distance between the table surface and flat detector - not more than 70 mm
  • Grid available
  • Tube motorized vertical movement available
  • Focus to floor distance - not less than from 350 to 2000 mm
  • Horizontal and vertical auto-tracking function available
  • Touch screen colour display interface for movement control - not less than 7.5 inch

X-ray stand

  • Vertical movement range (align top) - In the range of 400 to 2000 mm
  • Motorized vertical movement - not less than 1600 mm
  • Grid available
  • Vertical auto-tracking function available

X-ray radiation formation system

  • Maximum generator power - not less than 50 kW
  • Current adjustment range in radiography mode - not less than 10 – 600 mA
  • Voltage adjustment range in radiography mode - not less than 40 – 150 kV
  • Automatic exposure adjustment function available
  • X-ray tube focal spots number - not less than 2
  • Focal spot size - not more than 0.6 mm and 1.2 mm
  • Anode X-ray tube heating capacity - not less than 600 000 thermal units
  • Collimator with irradiation field illumination available
  • Dose meter (DAP) available

Two Flat Panel Detectors: in the Table (1) and in the stand (1)

  • Input field size - (43 × 43) ± 1 cm
  • Image Matrix - not less than 2800×2800
  • Pixel size - not more than 150 microns
  • DQE magnitude at 0 pairs of lines/cm - not less than 60%

Digital System of image processing

  • Saving images: hard disk capacity not less than 500 GB in mirror-image mode
  • Laboratory Assistant Workstation
  • Image postprocessing mode with the ability to:
  • Adjust brightness
  • Use compensational filters
  • Rotate images
  • Invert images
  • Increase (magnify)
  • Annotate
  • DICOM Network connection: Store, Print
  • Touch screen (sensor) remote control for exposure settings
  • LCD monitor with a diagonal of not less than 19": to view current images
  • Ability to record data reports on CD and DVD using DICOM file viewer program

Accessories and other equipment

  • Patient radiation exposure measuring device
  • Power distribution panel
  • UPS (Uninterrupted power supply) for digital system
  • Protective lead glass, non-framed - not less than 1000 x 900 mm

Maximum weight of the whole system (the installation premises can not sustain heavier weight) - not more than 1500kg