II.1.1) Title:
ÁTE - Acquisition of molecular genetic tools
Reference number: EKR001647512023

II.1.4) Short description: Lot No: 1- PCR device with daughter device

Part 1: PCR apparatus
# The main PCR device must have a universal block, which is suitable for handling 96 0.2 ml PCR tubes, 96-well plates or at least 48 0.5 ml PCR tubes.
# The gradient range of the device should be at least 1-20°C, across 12 columns.
# The device must have a flexible roof that ensures automatic roof adjustment when using plastic consumables of different heights.
# At least one daughter device must be connected to the device, which can be controlled from the base device.
# The temperature control range of the block is at least 4-99°C.
# The roof temperature of the device should be adjustable between 37-105°C.
# The block homogeneity between 20-72°C should not be higher than 0.3°C, at 90-95°C it should not be higher than 0.4°C. PCR daughter device:
# The daughter device should be a one-block PCR device without a programming interface that can be operated when connected to the main device. Programming from the base device to happen.
# The daughter device must have a universal block, which is suitable for handling 96 0.2 ml PCR tubes, 96-well plates or at least 48 0.5 ml PCR tubes.
# The device must have a flexible roof that ensures automatic roof adjustment
# when using plastic consumables of different heights.
# The temperature control range of the block is at least 4-99°C.
# The roof temperature of the device should be adjustable between 37-105°C.
# Block homogeneity between 20-72°C should not be greater than 0.3°C, at 90-95°C
# greater than 0.4°C.
# The heating speed should be at least: 2.5°C/s.
# The cooling speed should be at least: 2°C/s.
# The weight of the device should not exceed 11 kg.
# The heating speed should be at least: 2.5°C/s.
# The cooling speed should be at least: 2°C/s.
# The weight of the device should not exceed 11 kg.

II.2.7) Duration of the contract, framework agreement or dynamic purchasing system
Duration in months: 4

II.2.13) Information about European Union funds
Identification of the project:
RRF-2.3.1-21-2022-00001